How to write a technical specification for EEG, fNIRS, eye tracking and biosignal systems
How to write a technical specification for EEG, fNIRS, eye tracking and biosignal systems: vendor-neutral requirements, acceptance tests and common mistakes.
Published
A technical specification for research equipment is a document that defines the functions and performance a system must deliver as measurable, verifiable requirements, without pointing to a particular brand or model. A good specification gets you the device your research actually needs, lets more than one supplier bid, and settles from the outset what will be checked at delivery. This guide explains how to write a technical specification for EEG, fNIRS, eye tracking and biosignal systems. It is practical guidance, not legal advice; finalise your specification with your institution’s procurement unit and, where needed, its legal counsel.
Function instead of brand: measurable requirements
A specification starts from the research question, not from a product catalogue. First, write in plain sentences what the system has to do: which signals, in how many participants, in what environment, with which stimuli and alongside which other devices will they be recorded? Then turn each function into a requirement that carries a unit, a threshold and a measurement condition. Instead of “high-quality 64-channel EEG”, a line such as “at least 64 EEG channels recorded simultaneously, at least a stated sampling rate per channel, and an upper limit on noise with the bandwidth over which it is measured” makes clear both what the bidder must offer and what the committee will check.
A well-written requirement has four qualities:
- Measurable
- It is expressed as a number with a unit (channels, Hz, bits, µV, dB, nm, degrees) and does not rest on adjectives such as “high” or “advanced”.
- Verifiable
- It can be proven at delivery by a test or by the manufacturer’s technical documentation; a requirement that cannot be verified causes disputes during evaluation.
- Justified
- It is tied to a need of the research; asked why it is required, you can answer with a paradigm or an analysis method.
- Neutral
- It can be met by the products of more than one manufacturer; it does not contain a name, design or unusual value found in only one model.
What does Article 12 of Türkiye’s Public Procurement Law say?
For purchases made with public funds in Türkiye, the framework of a specification is set by Public Procurement Law No. 4734. Under Article 12 of the law, technical criteria “shall aim to ensure efficiency and functionality, shall not contain elements that hinder competition, and shall provide equal opportunity for all bidders” (our translation). The same article states that technical specifications shall also include provisions ensuring conformity with national and international technical standards where they exist, and it sets a clear rule: “No particular brand, model, patent, origin, source or product may be specified, and no features or descriptions aimed at a particular brand or model shall be included.”
The exception is narrow: where national or international technical standards do not exist, or where it is not possible to define the technical features, a brand or model may be named on condition that the words “or equivalent” are added. In practice this means that even if no brand is named, writing into the specification a registered name, a patented design or an unusual catalogue value found only in one manufacturer’s product can count as a description aimed at that model. In addition, some purchases for research projects are made under the exceptions in Article 3 of the law, with their own procedures; your institution’s procurement unit decides which procedure applies. Whatever the procedure, a function-based, neutral specification reduces both the risk of complaints and objections and the risk of buying the wrong device.
Typical specification lines by device class
The table below shows lines that often appear in specifications for research equipment and how each can be phrased measurably. You set the values in square brackets according to the requirements of your research; the table proposes no thresholds, it only shows how to phrase them.
| Device class | Line | Measurable phrasing |
|---|---|---|
| EEG | Channel count | At least [N] EEG channels and at least [M] auxiliary (AUX) channels recorded simultaneously |
| EEG | Sampling rate | Simultaneous sampling on all channels at no less than [X] Hz per channel |
| EEG | Resolution | Analogue-to-digital converter of at least [B] bits; input range and smallest step stated in µV or nV |
| EEG | Input noise | No more than [Y] µV; stating whether peak-to-peak or RMS and the bandwidth of measurement |
| EEG | Common-mode rejection (CMRR) | At least [Z] dB; stating the frequency at which it is measured |
| EEG | Trigger inputs | At least [K] bits of trigger input; triggers recorded in the same file as the EEG, at sample level |
| EEG | Raw data formats | Raw data exportable in an open format such as EDF, BDF or CSV, without an additional licence |
| EEG | Synchronisation | A common clock with the stimulus presentation software and other devices (such as TTL or Lab Streaming Layer); latency and jitter measured at delivery |
| fNIRS | Wavelengths | At least two wavelengths to separate oxygenated and deoxygenated haemoglobin; values stated in nm |
| fNIRS | Channel count and geometry | At least [N] sources, [M] detectors and [C] channels; source-detector distance and cap layout stated |
| fNIRS | Sampling rate | At least [X] Hz per channel |
| fNIRS | Short-separation channels | At least [S] short-separation channels to separate systemic signals from the scalp and skull |
| Eye tracking | System type | Screen-based or wearable (glasses) system; whether a chin rest is required |
| Eye tracking | Sampling rate | At least [X] Hz |
| Eye tracking | Accuracy and precision | Accuracy no worse than [A]° and precision (RMS) no worse than [P]° of visual angle; measurement condition stated |
| Eye tracking | Synchronisation and data | A common clock with EEG or the stimulus software; export of raw gaze data |
| Biosignals | Sensor types | Each required sensor listed, such as ECG, EMG, EDA, respiration and accelerometer |
| Biosignals | Simultaneous acquisition | Simultaneous recording of at least [N] sensors on one device with a common clock; sampling rate and resolution per channel |
| Biosignals | Connectivity | Wireless connection type and range; recording to the device’s internal memory where needed |
Keep two points in mind when using the table. First, every numerical value is meaningful only with a measurement condition: unless the bandwidth for noise or the distance and lighting for accuracy are stated, the claims of different manufacturers cannot be compared. Second, the sampling rate must be at least twice the highest frequency of interest (the Nyquist criterion), and in practice several times that is chosen; but asking for a high value merely because it is “better” can mean needlessly large files and more expensive devices. The recording details that the EEG publication guidelines (Keil et al., 2014) ask authors to report are also a good checklist of the values a specification should define.
In fNIRS, long channels in adults are typically placed at a source-detector distance of about 3 cm, while short-separation channels sit much closer and measure only the changes in the scalp and skull, which helps separate that signal from the brain signal. The fNIRS community’s publication guidelines (Yücel et al., 2021) offer a checklist for reporting method details such as optode geometry and signal processing; asking for the same details in the specification makes it easier to produce publishable data with the system you buy. In eye tracking, accuracy is the average offset of the measured gaze point from the true point, and precision is how consistent successive measurements are while looking at the same point (Holmqvist et al., 2012). Because manufacturers’ values are usually measured under ideal conditions, requiring the measurement in your own environment and with your own participants at acceptance gives a realistic comparison.
Scope of delivery: installation, training, support and warranty
The services that come with delivery should be written as measurably as the device itself. Otherwise, very different scopes may be offered for the same amount and the bids become impossible to compare.
- Installation
- Whether installation is done on site, whether integration with the experiment computer and stimulus software is in scope, and which test marks installation as complete.
- Training
- For how many people, how many hours or days, on which topics (preparation, recording, data export, basic analysis) and in which language training is given; whether a certificate is issued at the end.
- Technical support
- Support channels, response time, remote access and how support continues after the warranty period ends.
- Warranty
- Warranty period and scope (amplifier, caps, electrodes and software separately), where repairs are done and whether a replacement device is provided during repair.
- Software licences
- Whether the licence is perpetual or time-limited, how many users and computers it covers, and for how long updates are included.
- Documentation
- User manual, technical datasheet, electrical safety and conformity documents for devices connected to participants (CE marking and a declaration of conformity with standards such as IEC 60601-1), and a description of the file formats.
- Consumables and spare parts
- The cap sizes, quantity of electrodes and gel supplied with delivery; for how long spare parts will be available.
To make evaluation easier, ask bidders to answer every line of the specification item by item, citing the relevant page of the supporting document. This practice, common in research equipment specifications in Türkiye, lets the committee match each claim to a document.
Acceptance tests: what to verify at delivery
Acceptance tests are the concrete steps showing that the requirements of the specification are actually met at delivery, and the inspection and acceptance procedure relies on them. Writing the tests into the specification from the start prevents disputes at delivery. For a typical research system, the following steps can be considered:
- Inventory check: comparing every delivered item (amplifier, caps, electrodes, cables, licences) with the list in the specification.
- Noise recording: comparing the noise level in a short recording, with the inputs shorted or using the manufacturer’s test connection, against the stated value.
- Known-signal test: showing that a signal of given amplitude and frequency from a signal generator is measured correctly.
- Trigger and latency test: measuring the latency and jitter between trigger markers and the actual stimulus onset with a photodiode on the stimulus screen or with an audio signal.
- Impedance and preparation: demonstrating electrode preparation and impedance measurement on a participant.
- Data export: seeing that a recording can be exported in an open format and opened in independent software (such as EEGLAB or MNE-Python).
- Wireless connection and battery: verifying the stated range and battery life during a recording.
- fNIRS and eye tracking: showing that a signal is received from all fNIRS channels and the short-separation channels; for eye tracking, measuring accuracy after calibration in your own environment.
Open standards are a long-term safeguard for data formats. EDF is a simple and flexible format developed for exchanging and storing multichannel biological and physical signals; BDF is a 24-bit version of it. Requiring that recordings can be converted to community standards such as EEG-BIDS (Pernet et al., 2019) makes it easier for other researchers and joint projects to read the data years later.
Common mistakes
- Copying one manufacturer’s datasheet
- This is the most common mistake. Taking over a product’s catalogue values, registered names or a design unique to that product steers the specification towards a single model, may conflict with Article 12 and increases the risk of complaints. Derive the values from your research needs and check that more than one manufacturer can meet them.
- Unverifiable superlatives
- Phrases such as “highest quality”, “state of the art” or “best on the market” cannot be measured, cannot be tested at delivery and cause disputes during evaluation. Write every quality with a unit, a threshold and a measurement condition.
- Over-specifying
- Asking for a channel count, sampling rate or feature the research does not need narrows competition and raises the cost. For every line, ask: “Without this value, could I not answer my research question?”
- Under-specifying
- Forgetting trigger inputs, raw data access, software licences or training means additional purchases after delivery. In particular, state synchronisation with the stimulus presentation software explicitly.
- Giving no measurement condition
- Lines that state no condition for noise, accuracy or battery life lead to values measured in different ways being set side by side.
- Leaving services for later
- If installation, training and support are not defined in the specification, bids include them in different ways and comparison becomes hard.
Steps to follow when preparing a specification
- Write the research question and the experimental paradigms to be used in one paragraph.
- List the measurements each paradigm needs: channels and regions, sampling, triggers, accompanying devices, recording environment.
- Turn these needs into requirements with a unit, a threshold and a measurement condition.
- Check that the requirements are open to competition by comparing them with the technical documents of more than one manufacturer.
- Add the scope of delivery, the acceptance tests and the requirement for an item-by-item compliance response.
- Review the draft with your procurement unit and, where needed, with legal counsel.
This guide offers a practical framework and is no substitute for legal advice; the procedure to follow and the documents to require depend on your institution and on how the purchase is funded. For information and a quote on the EEG, fNIRS, eye tracking and biosignal systems in the catalogue, you can use the quote form.
Frequently asked questions
Can a brand be named in a technical specification?
Under Article 12 of Türkiye’s Public Procurement Law No. 4734, a technical specification may not name a particular brand, model, patent, origin, source or product, nor include descriptions aimed at a particular brand or model. Where national or international technical standards do not exist or the technical features cannot be defined, a brand or model may be named with the words “or equivalent”. For your own purchase, consult your institution’s procurement unit.
What should an EEG system technical specification include?
The core lines are channel count, sampling rate, analogue-to-digital resolution, input noise, common-mode rejection (CMRR), trigger inputs, raw data formats and synchronisation with other devices. Each line should carry a unit, a threshold and a measurement condition. Installation, training, support, warranty and acceptance tests should be defined as well.
How do you write a technical specification for an EMG system?
For research EMG, state measurably the number of muscles to record, the surface electrode type, the sampling rate and resolution per channel, the need for wireless connection and range, simultaneous recording with other sensors, and raw data export. Clinical EMG and nerve conduction devices are a separate class; do not mix a research specification with theirs.
What should an fNIRS technical specification cover?
State the wavelengths, the number of sources and detectors, the channel count and source-detector distance, the sampling rate per channel and the short-separation channels. Short-separation channels help separate systemic signals from the scalp and skull. If the system will be used with EEG or eye tracking, synchronisation should be a separate line too.
How should accuracy be written in an eye tracker specification?
Accuracy and precision should be written in degrees of visual angle, stating the conditions of measurement: distance, lighting, number of calibration points and whether a chin rest is used. The need for a screen-based or wearable system, the sampling rate and synchronisation with EEG should be defined as separate lines.
What happens in an acceptance test for research equipment?
Typical acceptance tests compare the delivered items with the list, record the noise level, measure a known signal, measure trigger latency with a photodiode, open the raw data in independent software and verify wireless range and battery life. Writing these tests into the specification from the start prevents disputes at delivery.
References and further reading
- Public Procurement Law No. 4734, Article 12 · Legislation Information System (in Turkish)
- Keil et al. (2014) · Publication guidelines and recommendations for studies using EEG and MEG, Psychophysiology
- Yücel et al. (2021) · Best practices for fNIRS publications, Neurophotonics
- Holmqvist, Nyström and Mulvey (2012) · Eye tracker data quality, ETRA
- Pernet et al. (2019) · EEG-BIDS, Scientific Data
- European Data Format (EDF) · An open file format for biological signals